CT Main Disconnect Panel

Bevco ISC designs and manufactures standard and custom disconnect panels for CT imaging systems. Our products are engineered to provide effective, cost-efficient solutions for any CT imaging application.
 
Bevco CT power disconnect panels provide emergency shutdown, remote electrical tripping, and overcurrent protection, and serve as local disconnects for these systems. They also meet OSHA Lock Out/Tag Out (LOTO) provisions. After a power loss, our panels restore power to CT equipment following a brief delay that allows incoming power to stabilize. This auto-restart function reduces the potential for damage to sensitive electronic devices caused by power fluctuations.
 
Our disconnect panels for CT imaging systems are compatible with all manufacturers’ equipment. Just provide the make and model of your system, and Bevco will provide a high performance, easy to install solution. Everything required for installation is included with each panel, including wiring diagrams for connecting incoming power, cooling system power, and emergency controls. Each panel is pre-configured for control and power signals.
 
Custom CT power disconnect panels are also available, designed and built in-house to meet your unique specifications.
 
Features & Advantages:
High quality components and designs eliminate guesswork
Installation time is reduced with single-point power connection
Optional surge suppressor feature to protect sensitive electronics during instances of poor power quality
Optional power monitoring feature with email & text message alert capabilities
Standard designs speed CT disconnect approval and manufacturing, and eliminate installation delays
Includes 2 remote emergency off-wall mounted pushbuttons
Standard configurations are suitable for use in 25kA short-circuit current rated power systems
High short-circuit current ratings and numerous custom configurations are available
For operator safety, all designs utilize low voltage circuit controls
Full capacity ground bars accommodate ground conductors sized as power conductors
Factory testing and standardized designs guarantee high product quality and reliability
Semi-flush or surface mounting
Technical support before, during, and after installation provided by Bevco’s experienced design team
UL and cUL listed, labeled, and tested; conform to applicable National Electric Codes
Optional room lighting controllers and other accessories also available
Made in the USA
 
Contact Bevco ISC to place your order or for more information on our CT power disconnect panels
 
Configuration:

Ordering Information:

Basic Panels

Basic

Part Number

3-Phase Voltage

Amperage

Nominal Dimensions

Height

Width

Depth

CT480-090-BB

440/460/480 50/60 Hz

90

24"

15"

9"

CT480-110-BB

440/460/480 50/60 Hz

110

30"

24"

9"

CT480-125-BB

440/460/480 50/60 Hz

125

30"

24"

9"

CT480-150-BB

440/460/480 50/60 Hz

150

30"

24"

9"

CT400-090-BB

380/400/415 50/60 Hz

90

24"

15"

9"

CT400-110-BB

380/400/415 50/60 Hz

110

30"

24"

9"

CT400-125-BB

380/400/415 50/60 Hz

125

30"

24"

9"

CT400-150-BB

380/400/415 50/60 Hz

150

30"

24"

9"

 

 

Option

Description

-BB

Basic Main Disconnect Panel

-BU

UPS Interface option.   Fully integrate emergency power off function if an Uninterruptible Power Supply (by others) is included in the system.   See below for compatible UPS Bypass Panel for UPS maintenance and safety.

 

 

System-matched UPS Bypass Panels

Basic Part Number

kVA Rating

Amperage @ 480V3PH

Nominal Dimensions

Height

Width

Depth

UB100-120

100 kVA

120

24"

15"

9"

UB150-180

150 kVA

180

24"

15"

9"

UP225-270

225 kVA

270

24"

15"

9"

UB300-360

300 kVA

360

30"

24"

9"

UB100-600

500 kVA

600

30"

24"

9"

 

 

Listed above are common options and configurations.  A wide variety of potential combinations of voltage, short-circuit current ratings, and amperage requirements can be accommodated. Please CONTACT US with your specific needs.